The Role of Preventive Maintenance in Medical Device Reliability
What does reliability mean for a medical device?
A reliable medical device performs to its specification whenever it is called on, for the whole of its service life. For a patient monitor, that includes accurate readings. For a defibrillator that can sit unused for weeks, it mostly means being ready on the one occasion it matters, with a charged battery and working pads.
The defibrillator example shows why reliability is harder to judge in healthcare than in many sectors. A device can power up and pass a quick glance while its calibration has drifted or its battery has lost capacity. The MHRA's Managing Medical Devices guidance makes this point about batteries, noting that maintaining battery function is particularly relevant for devices that are used infrequently, such as defibrillators.
For a service provider, reliability has two sides. Uptime is the obvious one. The less visible one is confidence that a device which appears to be working is actually performing correctly.
How does preventive maintenance improve medical device reliability?
Preventive maintenance for medical equipment improves reliability by finding and correcting the causes of failure before they reach a patient. A single planned visit usually combines several different jobs:
- Replacing parts with a limited life, such as batteries and seals, before they fail in use
- Checking accuracy against the manufacturer's specification and recalibrating where drift has crept in
- Inspecting cables, connectors, casings and load-bearing parts for wear that a user's pre-use check can miss
- Applying manufacturer software updates and any actions required by Field Safety Notices
- Recording what was found, so the next engineer and the customer can see how the device is ageing
MHRA guidance recommends performance testing against the manufacturer's specification both before and after maintenance. The "before" test is the one most easily skipped when a schedule is tight, yet it is the reading that shows how the device was performing in clinical use and whether its service interval is right.
The record matters as much as the work. MHRA expects maintenance records to show which version of a device is in use and whether it has been modified or repaired since supply, including software updates. Built up over several visits, that history is what turns individual services into reliability data. If you are building a programme from scratch, our guide to planned preventive maintenance covers the fundamentals.
What can't preventive maintenance do?
Being clear about the limits of medical device preventive maintenance helps set honest expectations with customers and stops reliability problems being misdiagnosed as poor workmanship.
Breakdowns will still happen
The MHRA's guidance acknowledges that breakdowns can occur even with comprehensive maintenance schedules. Some failures are effectively random, such as an electronic fault or accidental damage. Planned maintenance reduces failures caused by wear and drift, and a reliable service also needs fast corrective response, plus access to a loan or substitute unit for critical devices. For devices with good sensor data, it is worth understanding where predictive maintenance fits alongside it.
Servicing can introduce faults
Every intervention carries some risk. The MHRA's case examples on hoist incidents describe a care home where a hoist's legs collapsed during a transfer and the patient died. A locking knob had been overtightened, consistent with incorrect reassembly during or after servicing, and the fault was not detected before the hoist went back into use. The MHRA's position is that any serviced or repaired hoist must be functionally checked before it returns to service. Its general device guidance applies the same principle more widely, calling for performance testing, and electrical safety testing where applicable, before any device is brought back into use.
Age eventually outweighs maintenance
Planned maintenance extends a device's reliable life, though it cannot hold it there indefinitely. Parts become harder to source and manufacturers eventually withdraw support. When the government funded new radiotherapy machines for 28 trusts in May 2025, it estimated that replacing older machines would save as many as 13,000 appointments from being lost to equipment breakdown, with hospitals using machines over ten years old prioritised.
A service provider's most useful contribution here is evidence. A device history showing rising repair frequency gives the customer what it needs to plan replacement before reliability falls away. The MHRA's replacement criteria include reliability, checked against service history, and equipment downtime.
How should PPM intervals be set for medical devices?
PPM intervals should start from the manufacturer's instructions, then be adjusted for how heavily the device is used and the environment it works in. That is the approach set out in the MHRA's Managing Medical Devices guidance, which indicates that service intervals range from one month to two years depending on the device.
In practice, a few factors refine that starting point:
- Statutory examinations sit on top of PPM. Where regulations such as LOLER apply, MHRA guidance is clear that those tests are carried out in addition to the manufacturer's recommended maintenance. For patient hoists, the MHRA states that LOLER thorough examination is required at intervals of no more than six months.
- Usage changes the picture. MHRA notes that heavy use or irregular maintenance can shorten a device's life cycle, while limited use may extend it. Two identical pumps in different departments may need different treatment.
- Failure patterns shift with age. A 2025 study of multi-parameter patient monitors at a hospital in China, which modelled failure risk using expert judgement, found that the components most likely to cause failure differed between monitors in their first five years and those aged 11 to 15 years. A PPM routine designed when a fleet was new may be inspecting the wrong things a decade later.
- Risk should decide priority. When engineer capacity is stretched, life-supporting and high-risk devices such as ventilators and defibrillators need to be protected first.
Why are customers watching PPM compliance more closely?
Healthcare organisations that outsource maintenance remain responsible for making sure their devices are maintained appropriately. MHRA guidance places that responsibility on the healthcare organisation and notes that using an outside provider still requires in-house expertise to monitor and manage the contract.
Recent regulator action reinforces this. The MHRA's National Patient Safety Alert on patient hoists, issued in September 2026, lists inadequate or overdue maintenance and LOLER examination among the most common causes of fatal and serious falls. The MHRA received 22 reports of fatal incidents during patient transfers between 2015 and 2025. The alert tells organisations that outsource servicing to monitor compliance with the schedule, and requires device inventories to record when each hoist was last serviced and when its next service and LOLER examination are due.
Expect customers to ask for evidence of this. Figures worth being able to produce for each customer include:
- PPM completed on or before the due date, split by device risk category
- Breakdowns per device model over a rolling period, so recurring problems stand out
- Repeat call-outs on the same device shortly after a planned service
- Faults found and fixed during PPM, compared with faults that surfaced as breakdowns
- Devices approaching or beyond the manufacturer's stated service life
The regulatory side is covered in more detail in our guide to compliance requirements for medical equipment, and for the wider picture of servicing, inspection and lifecycle management, see our guide to looking after medical devices.
How do spare parts affect medical device reliability?
A service can be completed on time and still store up a future failure if the wrong part goes in. MHRA guidance says replacement parts should come from the manufacturer or be made to the same specification, and that pre-used parts should not normally be fitted. It also expects critical parts to be traceable back to their supplier, which matters when a manufacturer issues a recall because it lets you identify every device carrying the affected part.
Keeping up with manufacturer changes is the other half of this. MHRA guidance says whoever provides maintenance should be automatically alerted to changes in repair or maintenance instructions, including Field Safety Notices, so the work on site matches what the manufacturer currently requires.
Where does software fit in?
Most of the reliability gains described above depend on coordination as much as engineering skill. For an external service provider, field service management software keeps each device's schedule and history attached to the engineer's job. Every device is tracked individually and its PPM and calibration work is scheduled automatically. Before starting, the engineer on site can see the device's service history. Test results and calibration certificates are stored digitally against the device, which is what makes the compliance figures above practical to produce.
In-house clinical engineering teams looking after their own organisation's equipment have different needs, centred on the asset register and internal reporting, which is where a CMMS fits.
FAQs
Does preventive maintenance stop medical devices from failing?
It reduces failures caused by wear and drift, and it finds many problems before they reach patients. Random failures still happen, so a reliable service also needs fast corrective response and a plan for replacing devices as they age.
What is the difference between a pre-use check and preventive maintenance?
A pre-use check is a quick inspection by the user before each use to catch obvious faults. Preventive maintenance is scheduled work by a trained engineer, covering inspection, performance testing, calibration and part replacement. The MHRA's 2026 hoist alert states that pre-use checks are not a substitute for planned maintenance and statutory examinations.
What should a PPM record for a medical device include?
The device's unique identifier, the work carried out, parts fitted and their source, test results before and after maintenance, any software changes and the next due date. MHRA guidance also recommends a dated "serviced" label on the device showing when the next service is due.
Reliability you can evidence
Preventive maintenance is the foundation of medical device reliability, and it works best when intervals reflect how each device is really used and every serviced device is checked before it goes back into clinical use. The service history that builds up along the way is just as valuable, because it is what lets your customers see their equipment is safe and plan for the point where maintenance alone is no longer enough.
Service Geeni's medical device service management software is built around the asset, so schedules, test records, parts and engineer activity all sit against the device they relate to.