How to Ensure Compliance in Medical Equipment Maintenance

Medical equipment compliance depends on two things: maintaining each device the way the manufacturer and the law require, and being able to prove it. For service providers, the legal duty usually sits with the healthcare customer, but that customer relies on your engineers' work and records to show its equipment is safe. This guide explains which UK rules apply and how to run a maintenance service that stands up to an audit.

What does medical equipment compliance mean?

In maintenance terms, medical equipment compliance means keeping devices in service safe and performing to specification, in line with the manufacturer's instructions and any legislation that covers that type of equipment, with records showing the work was done.

It helps to separate this from the regulatory compliance manufacturers deal with. UKCA marking and the Medical Devices Regulations 2002 mostly govern how devices are designed and placed on the market. Maintenance compliance starts once a device is in use, and it is judged on the condition of the equipment and the evidence held about it.

That evidence matters as much as the work. The MHRA's Managing Medical Devices guidance points out that a negligence defence based on good device management can only succeed if records are available for the device involved.

‍

Which regulations apply to medical equipment maintenance in the UK?

No single law covers medical equipment maintenance, and which rules apply depends on the device and the setting. The main ones are:

  • Regulation 15 of the Health and Social Care Act 2008 (Regulated Activities) Regulations 2014. In England, CQC-registered providers must keep equipment properly maintained. CQC's guidance on Regulation 15 says maintenance arrangements must meet current legislation, guidance and manufacturers' instructions.
  • The Health and Safety at Work etc. Act 1974 and the Electricity at Work Regulations 1989. The MHRA identifies these as the basis for electrical safety testing programmes in healthcare.
  • The Provision and Use of Work Equipment Regulations 1998 (PUWER). Equipment used by staff at work must be maintained in an efficient state, in efficient working order and in good repair.
  • The Lifting Operations and Lifting Equipment Regulations 1998 (LOLER). Patient hoists and slings need a thorough examination by a competent person at intervals of no more than six months.
  • The Ionising Radiations Regulations 2017. X-ray and other radiation equipment has its own testing requirements, and contracts for it should be drawn up with advice from a radiation protection adviser.
  • The Medical Devices Regulations 2002. These mainly apply to manufacturers but can reach maintainers. MHRA guidance warns that modifying a device, or using it for a purpose the manufacturer didn't intend, may count as manufacturing a new device, and a fully refurbished device supplied to a new owner needs a new UKCA or CE mark from the refurbisher.

The MHRA's Managing Medical Devices guidance sits across all of these. It is guidance rather than law, but it is written to help healthcare organisations meet Regulation 15 and the MHRA still points organisations to it in current safety alerts. It also makes clear that statutory tests under LOLER or the Ionising Radiations Regulations are carried out on top of the manufacturer's recommended maintenance and never replace it.

Regulation 15 only applies in England. Scotland, Wales and Northern Ireland have their own care regulators and guidance, such as Scottish Health Technical Note 00-04, so check which framework covers each customer site. For the wider picture of servicing, calibration and lifecycle management, see our guide to maintaining healthcare equipment.

‍

Who is responsible for compliance when maintenance is outsourced?

The healthcare organisation stays legally responsible for its equipment, even when an outside company carries out the maintenance. CQC states that providers keep their legal responsibility under Regulation 15 when they delegate work through contracts to third parties. The MHRA tells healthcare organisations that using an external provider still requires in-house expertise to monitor and manage the contract.

Service providers still carry obligations of their own:

  • Engineer competence. Health and safety law requires employers to train staff adequately, and MHRA guidance expects maintainers to hold written evidence of up-to-date training.
  • Liability and insurance. The MHRA warns that if a device fails after a part not matching the manufacturer's specification has been fitted, and a patient or user is seriously injured or killed, the organisation responsible for the repair is more likely to be held liable. It also tells healthcare organisations to confirm their maintenance provider has adequate insurance or indemnity.

Customers are also being told to check their providers more closely. The MHRA's National Patient Safety Alert on patient hoists, issued in September 2026, is addressed to equipment providers as well as care organisations. It lists inadequate or overdue maintenance and LOLER examination among the most common causes of fatal and serious falls. Organisations that outsource servicing are told to monitor compliance with the schedule and to take any hoist with an overdue examination or unresolved defect out of service.

‍

How do you ensure compliance in medical equipment maintenance?

For a service provider, compliance is easiest to sustain when the requirements are built into how every job is planned and closed. The steps below cover the main areas to get right.

Agree responsibilities in the contract

The service contract is where shared responsibility gets defined. MHRA guidance says a maintenance contract should cover, where relevant:

  • Reference to the manufacturer's written instructions
  • Availability, source and traceability of spare parts
  • Notification of any changes, including alternative parts or methods
  • Staff training and quality management systems
  • Record keeping requirements
  • Response times and loan devices
  • Use of subcontractors

If any work is subcontracted, the customer should know about it, and the MHRA expects the main provider to audit the subcontractor regularly.

Build schedules from the manufacturer's instructions

The frequency and content of planned maintenance should start from the manufacturer's instructions, adjusted for how heavily the device is used and the environment it works in. Statutory examinations such as LOLER, along with electrical safety testing, are then scheduled alongside. We cover interval setting in more detail in our post on how preventive maintenance supports device reliability.

Statutory examinations also depend on every item being identifiable. HSE's guidance on moving and handling equipment explains that under LOLER, anyone using lifting equipment should be able to check it has been thoroughly examined, so hoists and slings need to be individually identifiable.

Match jobs to trained engineers

MHRA guidance states that anyone without adequate training should not attempt to repair or maintain a medical device. Job allocation therefore needs to account for which engineers are trained on which models.

Act on Field Safety Notices

A Field Safety Notice (FSN) is a manufacturer's notice telling users about safety action needed on a device, such as a modification or a software update. The MHRA states that anyone who receives an FSN must always act on it, and its weekly Field Safety Notice lists show the volume involved: 13 notices were listed for the last week of October 2026 alone.

MHRA guidance says maintenance providers should be automatically alerted to changes in maintenance instructions, and that records should show the version of each device in use, including its software. Recording which FSN actions are complete on which devices lets the customer show each notice was closed out.

Test before return to service

Before a device goes back into use after maintenance or repair, MHRA guidance says it should be tested for performance and, where applicable, electrical safety. The user should be told the results and any changes made to the device's settings. For electrical safety testing, the MHRA recommends BS EN 62353, the UK version of IEC 62353, as best practice. The test equipment itself also needs regular calibration.

Handle incidents and modifications correctly

When a device has been involved in an incident, the MHRA's Yellow Card guidance says it should not be thrown away or repaired. An engineer attending a fault that has harmed a patient should leave the device unrepaired, keep it out of use and report through the customer's procedures, usually via its Medical Device Safety Officer (MDSO) where it has one. Incidents in England, Wales and Northern Ireland go to the Yellow Card scheme, while Scotland uses the Health Facilities Scotland reporting system.

Modifications need similar care. Any change outside the manufacturer's intended use should only happen as part of a fully documented risk management process, because it can shift liability onto whoever made the change.

‍

What records prove medical equipment compliance?

Good compliance records let anyone pick a device and see its complete history. According to MHRA guidance, device records should provide evidence of:

  • A unique identifier for the device
  • A full history, including when it was put into use and where it is deployed
  • Any specific legal requirements and whether they have been met
  • The schedule and details of maintenance and repairs
  • The end-of-life date, where one is specified

At job level, that means capturing the work done, parts fitted and where they came from, test results, software versions, FSN actions, the engineer who did the work and the next due date. MHRA guidance also says records must be protected so that later changes do not obscure what was recorded before, and must be available for inspection by bodies such as CQC.

Healthcare organisations are told to carry out random audits of maintenance work and the records behind it, so any device you look after could be selected. Reviewing overdue PPM and statutory examinations by customer, and checking open FSN actions against the devices they affect, reduces the chance of a gap surfacing at audit.

‍

How does software support medical equipment compliance?

Most of the steps above depend on information being linked to the right device. For external service providers, field service management software keeps each device's schedule and history attached to the engineer's job. PPM is scheduled automatically, calibration certificates and test results are stored against the device, jobs are matched to engineers by skill set and parts are tracked by serial number. Through a customer portal, customers can download service reports and view compliance documents for themselves.

In-house clinical engineering teams maintaining their own organisation's equipment have similar record-keeping needs, usually met by a CMMS, which our guide to CMMS and compliance reporting covers.

‍

Medical equipment compliance FAQs

Who regulates medical equipment maintenance in the UK?

The MHRA regulates medical devices and publishes guidance on managing them. In England, CQC inspects registered providers against Regulation 15 and is the relevant regulator for patient safety in those services, while health and safety law such as LOLER and PUWER also applies.

Is IEC 62353 testing a legal requirement?

No. MHRA guidance notes there is no specific legal obligation to carry out any particular electrical safety test, although there is a general duty to protect staff and patients from danger. The MHRA recommends BS EN 62353, the UK version of IEC 62353, as best practice for recurrent testing and testing after repair.

Do medical equipment service providers need ISO 13485?

MHRA guidance does not make it a requirement. It recommends that healthcare organisations use a service provider that complies with a relevant quality system standard, such as ISO 13485 or ISO 9001, where one is available, so certification can count in your favour when customers choose a provider.

‍

Making compliance part of every job

Medical equipment compliance rests with your customer in law, but much of the evidence behind it comes from your service. Building the requirements into everyday jobs means that evidence exists for every device and protects your business if something goes wrong.

Service Geeni's medical device management software is built around the individual asset, so schedules, test records, parts and engineer activity all sit against the device they relate to.

Download Full PDF

Enter your email address and download our full guide today.
Download Now
Thank you! Your submission has been received!
Download PDF
Oops! Something went wrong while submitting the form.

Table of contents

Book Your Demo Today

Get a personalised walkthrough of Service Geeni, tailored to your business needs. See how our asset-centric, industry-specific platform can streamline your operations, reduce downtime, and help your team work smarter.

Talk to Us

Have questions or want to know how Service Geeni fits into your business? Whether you’re looking for integrations, pricing, or just some expert advice, we’re here to help. Reach out to our team and let’s chat about your needs.